High resolution quantitation of microcystins and nodularin on SCIEX X500R QTOF System


KC Hyland, April Quinn-Paquet, Craig Butt
SCIEX, USA

Abstract

Toxic contaminants in drinking water include microcystins and nodularins which are regulated under the US EPA UCMR4 list. While triple quadrupole systems are typically used for quantitation because of their sensitivity in robustness, the ability to obtain confirmatory information has made high resolution/accurate mass (HRAM) systems important platforms for drinking water analysis. Here an MRMHR workflow has been developed to acquire both MS and MS/MS data on the SCIEX X500R QTOF System for quantitation and confirmation of suite of eight microcystins and one nodularin, reaching the MRLs defined in the UCMR4 list in drinking water matrices. Sensitivity was demonstrated by LOQ values ranging from <0.1μg/L - 3μg/L.


Introduction

Toxic products of cyanobacteria and aquatic microfaunal communities include the implicated human carcinogens microcystins and nodularins. Microcystins represent an important contaminant class for environmental and drinking water analysis under the US EPA UCMR4 list, effective starting in 2018.

Analysis of microcystins and many other water contaminants is commonly conducted on triple quadrupole systems for sensitivity and robustness. Environmental testing labs are frequently confronted with the need for more confirmatory information when it comes to analyte identification in real world samples. To this end, there has been increased interest and adoption of high resolution/accurate mass (HRAM) systems. Advantages gained include the ability to garner more specificity and qualitative confirmation of compound identification, while maintaining the capacity for sensitive and robust, routine quantitation (Figure 1).

Methods and data are presented which demonstrate use of HRAM on a QTOF mass spectrometer for accurate and sensitive quantitation of microcystins and nodularin in water samples.

Figure 1. Target identification points of confirmation.  Identification and quantitation of microcystins in unknown samples can be achieved with high confidence using HRAM analysis due to multiple points of matching using accurate mass, MRMHR, isotope pattern matching, ion ratio, and chromatographic retention time.

Key assay attributes

  • Quantitation of a suite of eight microcystins (MC) and one nodularin (NOD) was achieved using high resolution accurate mass LC-MS/MS with the SCIEX X500R QTOF System.
  • Utilizing data acquisition with high resolution Multiple Reaction Monitoring (MRMHR workflow) in conjunction with the simultaneous collection of TOF MS data, multiple approaches to achieving accurate and sensitive quantitative analyses for these analytes in drinking water matrices are presented.
  • Sensitivity was demonstrated by LOQ values ranging from <0.1μg/L - 3μg/L.
  •  US EPA 10-day drinking water health advisory for microcystins is 0.3 μg/L for infants and children up to 6 years old, and 1.6 μg/L for adults.
  • Confidence in compound identification is imparted by the specificity of accurate mass data for the ionized compound in TOF MS mode, as well as the product ion generated from the isolated precursor in MRMHR workflow (Figure 1).

 

Experimental

Standards: Neat standards were obtained from Enzo Life Sciences (Farmingdale, NY) and reconstituted in 1 mL of methanol. An intermediate mixed stock was prepared by diluting the standards in methanol to yield 500 μg/L for MC-RR and Nodularin-R, and 2000 μg/L for MC-LA, MC-LF, MC-LR, MC-LY, MC-LW, MC-YR, MC-WR. Calibration standards were prepared with 5% acetonitrile in water to match the initial LC conditions. Standards were prepared in glass vials and kept at -20°C until analysis.

HPLC system: A SCIEX ExionLC™ AD System was used as the LC system. Chromatographic separation was achieved under gradient conditions using a Phenomenex Kinetex® C8 column (2.6 μm particle size, 100 x 3 mm) and flow rate of 0.500 mL/min. The mobile phases were water with 0.1% formic acid (“A”) and acetonitrile with 0.1% formic acid (“B”). Column oven temperature was 40°C and a 20 μL injection was used. To reduce sample carryover the autosampler rinse solvent was 60:20:20 isopropyl alcohol: methanol: acetonitrile using a rinse volume of 2 mL and dip time of 8 s. Run time was 11 minutes for the full gradient. An example elution profile is shown in Figure 2.

Figure 2. Chromatographic separation of 8 microcystins and 1 nodularin. Separation was achieved using a SCIEX ExionLC AD system and a Phenomenex Kinetex C8 column over an 11-minute gradient with water and acetonitrile.

Mass spectrometry: The SCIEX X500R QTOF System with the Turbo V™ Ion Source was operated in positive mode electrospray ionization (ESI). Source parameters are listed in Table 1. The TOF MS scan was conducted over a range of 500 to 1100 m/z. MRMHR experiment included two transitions monitored for each analyte. For each transition in the acquisition method, the precursor ion was defined for the target analyte, and a mass range was defined which would encompass the expected fragment ion. Optimized declustering potential (DP) and collision energy (CE) voltages were designated for each transition, based on optimized values determined from the Guided MRMHR algorithm. Additionally, retention time (RT) values were specified for each MRMHR transition, with RT tolerance values of 15 s for each.

Table 1. Ion source parameters. Electrospray Ionization (ESI) was conducted in positive ion mode.

Data analysis and method performance evaluation: Data acquisition and processing were performed using SCIEX OS Software. Calibration curves were generated from the injection of seven concentration levels from 0.12 to 312 µg/L. Linear regression models with 1/x weighting were used to fit all calibration curves, and 1-point Gaussian smooth was applied to data for calibration and sensitivity evaluation. Example calibration curves for a subset of analytes can be seen in Figure 3. For the evaluation of method sensitivity, the limit of detection (LOD) was defined as the lowest calibrator with peak-to-peak signal to noise ≥3. The limit of quantitation (LOQ) was defined as the lowest calibrator with peak-to-peak signal to noise ≥10.

Table 2. Accurate masses for microcystins and nodularin. The formula and accurate monoisotopic m/z for each analyte studied is provided.

Key quantitation limits for assay

US EPA 10-day drinking water health advisory for microcystins is 0.3 μg/L for infants and children up to 6 years old, and 1.6 μg/L for adults. Maximum acceptable concentration (MAC) of MC-LR is 1.5 μg/L in Canada, and the World Health Organization (WHO) MC-LR provisional guideline is 1 μg/L. The EPA additionally has included several of these compounds of its Fourth Unregulated Contaminant Monitoring Rule (UCMR4). Minimum Reporting Levels (MRLs) ranging from 0.005 to 0.09 μg/L are listed in this document based on EPA standard Method 544. EPA Method 544 utilizes LC-MS/MS with up front concentration of water samples using solid phase extraction (SPE).

Quantitation results

Quantitation was achieved by utilizing high resolution TOF MS data of each precursor. MRMHR transitions were acquired and used for quantitation and confirmation. Both scans occur within every cycle during an injection, and processing of the acquired data can utilize either or both. Comparison of these two quantitative analyses indicate that quantitation of these microcystins using the TOF MS extracted accurate mass is sensitive, linear, and precise; and demonstrates better performance compared to MRMHR for quantitation. TOF MS data was processed using known analyte charged monoisotopic masses (m/z) and a mass extraction width (XIC width) of 0.02 Da. Table 2 shows the extracted precursor masses used. Microcystins in the analyzed suite have MW >900. Ionization was performed in positive ion mode, but it is noted that the target MCs will protonate to form [M+H]+ species as well as doubly charged species. Eight of the nine analyzed targets show the singly charged species as dominant; it was therefore used for monitoring and quantitation. However, MC-RR showed the +2 charge as a greater intensity precursor than [M+H]+. The m/z values monitored for this precursor reflect the choice of the [M+2H]++ species for maximized sensitivity (Table 2).

The 11-minute chromatographic gradient with the Kinetex C8 column achieved baseline separation for all compounds (Figure 2). The gradient is 15 min shorter than the program described in EPA Method 544, resulting in considerable time savings while maintaining separation and peak quality.

The LOD and LOQ concentrations for all compounds were demonstrated to be well below the lower range of the calibration curve (lowest calibrator 0.12 μg/L) when analyzing using TOF MS data (Table 3). Signal- to- Noise ratios at 0.12 µg/L ranged from 27 (for MC-LR) to 91 (for MC-RR) demonstrating that lower concentrations may be detected and quantified using this method. Precision is shown as the %CV of triplicate analyses and is also listed in Table 4 for two concentration levels. Reproducibility is excellent with all %CV values of 7% or less for the TOF MS data. Linear dynamic range was also calculated and shown to be between 2.5 and 3.4 orders of magnitude. MRMHR data showed LOQ concentrations in the range of 1 – 3 μg/L.  Precision at the low levels is also reduced compared to the TOF data, likely as a result of the decrease in the absolute signal available in the MS/MS data.

A key advantage of HRAM approach to quantitation is the ability to apply multiple points of confirmation to avoid reporting false positive detection (Figure 1). The five points of confirmation applied in this method are:

  1. Accurate mass of the target precursor
  2. Confirmation with the MRMHR transition and the accurate mass of the fragment ion
  3. Isotope ratio matching to theoretical pattern based on target formula
  4. Ion ratio matching between standard and unknown sample
  5. Retention time matching between standard and sample

Figure 3. Calibration curves for MC-LR. Curves are shown for one of the target analytes, MC-LR, for three different acquisition channels. The isolation of the TOF MS accurate mass (top), as well as the two optimized MRMHR transitions, can all be detected in calibration standards to produce a regression for quantitation.

Table 3.  Quantitative analysis method performance. Detection and quantitation limits, linearity, and reproducibility for the microcystins suite. TOF MS data processed with a mass extraction width (XIC width) of 0.02 Da.

High MS data quality for confirmation

High resolution TOF MS analysis of microcystins shows mass resolution >40,000 across the range of precursor m/z. This mass resolution infers a high degree of specificity for the target analyte. Mass accuracy within 2ppm reinforces confidence in the identification of the targets. High resolution fragments collected using MRMHR also demonstrate excellent specificity and confidence from high mass resolution and mass accuracy (Figure 4 and 5). TOF MS pattern of resolved MS isotopes was compared to the theoretical isotope pattern for the empirical formulae of each microcystin; these data provide additional confirmation of the presence of the compound and are displayed in the results table within the SCIEX OS Software. Precursor XIC and MRMHR transition data can also be grouped together in the data processing module of SCIEX OS Software, so that peak data corresponding to the same compound are processed as such; this allows the software to automatically evaluate ion ratios by comparing signal from multiple acquisition channels for a single analyte. Ion ratio comparison of sample data to that of standards provides another metric for qualitative matching and identity confirmation.

Figure 4. HRAM TOF MS analysis of microcystin MC-RR.  Mass spectral data shows high mass resolution and mass accuracy,  providing specificity for the target analyte. Mass accuracy within 2ppm for the precursor for the TOF MS precursor reinforces confidence in the identification of the targets. MRMHR fragments also demonstrate excellent specificity and confidence with resolution >28,000 and mass error <6ppm.

Figure 5. HRAM TOF MS analysis of microcystin MC-YR and nodularin. Mass spectral datashows high mass resolution and mass accuracy, providing specificity for thesetarget analytes. Mass accuracy within 2ppm for the precursor for the TOF MS precursor reinforces confidence in the identification of the targets. MRMHRfragments also demonstrate excellent specificity and confidence with resolution >28,000 and mass error <5ppm

Summary

Quantitation of a suite of microcystins and nodularin was demonstrated using LC-MS/MS on SCIEX X500R QTOF System. Utilizing the TOF MS spectral data and extracting a narrow XIC for analyte precursor ions provided excellent quantitation results with sensitive detection (<0.1μg/L LOD) and exceptional reproducibility (%CV <7%) at low concentrations. If used with the EPA Method 544 suggested concentration factor of 500-fold, these detection limits are theoretically extended to 0.0002 μg/L, Concurrent MRMHR data allows for additional quantitative information and enhanced confidence in qualitative analysis without requiring additional injections or processing. Using the HRAM approach on the X500R system, confirmation of target identification can be attained using accurate precursor mass matching, isotope pattern matching, accurate fragment mass matching, ion ratio matching, and retention time matching. Verifying across these five points reinforces the ability to confirm the presence of the analyte and helps safeguard against reporting false positives.

 

Related content

  1. Quantitation of Microcystins and Nodularins in Water Samples Using LC-MS/MS. SCIEX technical note RUO-MKT-02-5272-A.