Abstract
A study was performed on the SCIEX 7500 System to assess the robustness for bioanalysis on a very large sample set of 1500 injections. Four compounds were analyzed per injection at concentrations of 10 ng/mL in plasma. Raw peak area reproducibility across the 1500 injections was 3.12 – 5.02%.
Introduction
Here, the robustness of the SCIEX Triple Quad 7500 LC-MS/MS System — QTRAP Ready was evaluated over 1500 injections of a plasma sample spiked with four relevant bioanalysis compounds (atenolol, metoprolol, nadolol and rosuvastatin). This number of injections relates to more than two days total analysis time and shows that the instrument can perform optimally even after this large number of injections. After this testing, no instrument cleaning was required before the next analysis could begin, showing that the system’s performance had not been impacted.
Figure 1, below, plots area values of each of the analytes against injection number, with a successful outcome appearing flat across the injection number. For this analysis, internal standards have not been used, so the data shown represents uncorrected, raw areas for each analyte. This highlights the robustness of the system.
More details on the sample preparation and methods can be found in the Supplementary information.

Conclusions
- The SCIEX 7500 System provides excellent levels of robustness for bioanalysis from plasma matrices and provides consistent area values over more than two days of analytical run time
- Analysis was performed in both positive and negative mode, highlighting that the SCIEX 7500 System is capable of high levels of robustness in both acquisition modes
- No cleaning or maintenance of the instrument was necessary after this analysis was complete.
Acknowledgements
The samples and methods needed to perform this analysis were provided by Cyprotex Discovery, UK.