Sensitive analysis of the proteolysis targeting chimera (PROTAC) degrader ARV-110 in rat plasma


Enabling sensitive quantification of PROTAC degraders in matrix using the SCIEX 7500 system

Junmiao Chen1 , Dandan Si1 , Long Zhimin1 , Eshani Nandita2 and Rahul Baghla2
1SCIEX, China; 2SCIEX, USA

Abstract


This technical note demonstrates the sensitive quantification of the PROTAC degrader ARV-110 (bavdegalutamide) in rat plasma using a high-end triple quadrupole mass spectrometer. Low-pg/mL quantification was achieved with outstanding accuracy, precision and linearity. The improved front-end technology on the SCIEX 7500 system enabled greater ion generation, capture and transmission for enhanced sensitivity for bioanalysis of PROTACs in complex matrices.

Introduction


The concept for employing PROTACs for targeted protein degradation was first demonstrated in 2001 by Sakamoto et al.1 Since then, considerable developments have been made to apply this approach to treat drug-resistant targets. 2 PROTACs are heterobifunctional compounds composed of a ligand for a protein of interest (POI) and an E3 ligase that are connected through a linker (Figure 1). PROTACs initiate degradation by forming a ternary complex with the POI followed by ubiquitination. The ubiquitinated POI is then detected and degraded by endogenous 26S proteasomes present in eukaryotic cells.2 The unique chemical knockdown strategy of PROTACs makes it a remarkable technology for targeted protein degradation. Additionally, PROTACs reach complete efficacy at low dosages compared to conventional small molecules. Therefore, highly sensitive bioanalytical techniques are needed to perform pharmacokinetic and pharmacodynamic studies to evaluate product safety and efficacy.

Here, ARV-110 was employed as a model PROTAC to develop a quantification method in rat plasma using the SCIEX 7500 system. The improvement in the front-end technology of the SCIEX 7500 system enabled a lower limit of quantification (LLOQ) of 5 pg/mL to be achieved for ARV-110 using a protein precipitation sample preparation. The assay covered a calibration range covering 5 pg/mL to 5000 pg/mL with a linear dynamic range (LDR) spanning 3 orders of magnitude.

Key features of bioanalysis of PROTACs in matrix using the SCIEX 7500 system
 

  • Sensitive quantification: Achieved a low-pg/mL level LLOQ for the quantification of ARV-110 in rat plasma with minimal sample preparation

  • Excellent quantitative performance: Achieve outstanding accuracy, precision and linearity for the analysis of PROTACs using the SCIEX 7500 system featuring improved front-end technology

  • Streamlined data management: Increase productivity with a user-friendly interface and integrated platform for data acquisition, processing and management for routine bioanalysis using SCIEX OS software

Figure 1. Structure of ARV-110. The PROTAC consists of a ligand for a POI connected to an E3 ubiquitin ligase through a linker.

Methods

Sample preparation: ARV-110 was spiked into 100 μL rat plasma at concentrations ranging from 5 pg/mL to 5000 pg/mL. Protein precipitation was performed using a 600 μL aliquot of 1:1 (v/v) methanol:acetonitrile. Samples were vortexed for 1 min followed by centrifugation at 13,000 rpm for 10 min at 4°C. The supernatant was aliquoted and dried under nitrogen gas. Samples were reconstituted using 200 μL of 1:1 (v/v) methanol:water prior to analysis.

Chromatography: Samples were analyzed using an ExionLC AC system at a flow rate of 0.3 mL/min with a Phenomenex Kinetex C18 (2.1 x 100 mm, 2.6 µm, 100 Å) column and a 10 minute gradient (Table 1). The operating column temperature was 40°C. Mobile phase A was 0.05% formic acid in water and mobile phase B was 0.05% formic acid in acetonitrile. A 10 µL injection volume was used for analysis.

Table 1. Chromatographic gradient.

Mass spectrometry: Samples were analyzed using the SCIEX 7500 system in positive MRM mode. The MRM conditions and optimized MS parameters are listed in Tables 2 and 3, respectively.

Table 2. MRM parameters and fragments used for quantification. 

Table 3. MS parameters. 

Data processing: Data processing was performed using the Analytics module of SCIEX OS software. A weighting of 1/x2 was used for quantification.

Figure 2. Calibration curves for ARV-110 in rat plasma. A weighting of 1/x2 was applied for quantification. Strong linearity was achieved with a correlation coefficient (r2) of 0.996.

Method development and quantitative performance


Several PROTACs have been validated at the preclinical stage including ARV-110, which reached clinical trials in 2019.3 With several promising PROTAC candidates in the horizon, thorough preclinical and clinical analyses must be performed. To support proper drug product evaluation for safety and efficacy, rapid and sensitive bioanalytical methods are needed for an accurate quantitative determination in matrix.

In this study, the SCIEX 7500 system was employed for the quantification of ARV-110 in rat plasma. The SCIEX 7500 system is equipped with key hardware features that provide significant gains in the generation, capture and transmission of ions. This offers high levels of sensitivity and quantification power for analytes in complex matrices. 

A [M+H]+ at m/z 812.4 was selected as the precursor ion of ARV110. The most dominant fragment ion of ARV-110 was at m/z 452.2 and was chosen for quantification. For structural confirmation, a secondary fragment ion at m/z 562.2 was monitored.

PROTAC concentrations ranging from 5 pg/mL to 5000 pg/mL were analyzed and a LDR spanning 3 orders of magnitude was observed. Strong linearity was reached with a correlation coefficient (r2 ) of 0.996 (Figure 2).

An LLOQ of 5 pg/mL was achieved for ARV-110 in rat plasma with a total run time of 7 mins (Figure 3). No interferences were observed in the matrix blank at the retention time of the compounds, as shown in Figure 3.

The quantification results are summarized in Table 4. Overall, the accuracy was within ±11.2% of the nominal concentration for ARV-110.

Figure 3. Representative extracted ion chromatograms (XICs) for ARV-110 in rat plasma. An LLOQ of 5 pg/mL was achieved for ARV-110. No matrix interferences were observed in the blank.

Table 4. Accuracy achieved at each concentration level measured.

Conclusion
 

  • An LLOQ of 5 pg/mL was reached for quantification of ARV110 in rat plasma with minimal sample preparation

  • A highly sensitive assay for quantification of PROTACs was demonstrated on the SCIEX 7500 system with an improved front-end technology for better ion generation, capture and transmission

  • The method demonstrated excellent accuracy, precision and linearity at all concentration levels

  • A single platform for streamlined data acquisition, processing and management with SCIEX OS software is presented 

References
 

  1. Kathleen M. Sakamoto, Kyung B. Kim, Akiko Kumagai, Frank Mercurio, Craig M. Crews and Raymond J. Deshaies (2001). Protacs: Chimeric molecules that target proteins to the Skp1–Cullin–F box complex for ubiquitination and degradation. PNAS 98(15): 8554-8559.

  2. Mariell Pettersson and Craig M Crews (2009). PROteolysis TArgeting Chimeras (PROTACs) - Past, present and future. Drug Discov Today Technol. 31: 15-27.

  3. Thi-Thao-Linh Nguyen, Jin Woo Kim, Hae-In Choi, Han-Joo Maeng and Tae-Sung Koo (2022). Development of an LCMS/MS Method for ARV-110, a PROTAC Molecule, and Applications to Pharmacokinetic Studies. Molecules 27(6): 1977