Quantitative analysis of 9 steroid hormones in human serum
Abstract
This technical note demonstrates the accurate quantitation of 9 steroid hormones in human serum using a solid-phase extraction (SPE) procedure with analysis using the SCIEX QTRAP 4500 system. Low-nmol/L level sensitivity was achieved with excellent precision (ranging from 0.7 to 8.0%) and accuracy (ranging from 92.0 to 111%) at the lowest calibrator, demonstrating the quantitative performance of the assay. The signal-to-noise ratios (S/N) were 29:1 for corticosterone, 446:1 for cortisol, 55:1 for androstenedione, 35:1 for progesterone, 26:1 for 11-deoxycortisol, 16:1 for 21-deoxycortisol, 41:1 for 17-OH progesterone, 145:1 for testosterone and 1599:1 for DHEAS at the lowest calibrator measured. In addition, excellent linearity was observed across clinically relevant concentrations, with r2 values above 0.99 across the calibration ranges for all the analytes.
Key benefits of steroid hormone analysis from human serum using the QTRAP 4500 system
Efficient sample preparation: Steroid hormones were extracted from human serum samples using an efficient solid-phase extraction (SPE) procedure, requiring 100 μL of human serum
Excellent quantitative performance: Sensitive quantitation of steroid hormones was performed with excellent precision and accuracy (4.6% and 111% for corticosterone at 0.2 nmol/L, 3.9% and 104% for cortisol at 8.3 nmol/L, 2.4% and 108% for androstenedione 0.5 nmol/L, 0.7% and 110% for progesterone at 0.4 nmol/L, 8.0% and 107% for 11-deoxycortisol at 0.4 nmol/L, 1.2% and 92.0% for 21-deoxycortisol at 0.3 nmol/L, 7.4% and 98.6% for 17-OH progesterone at 0.3 nmol/L, 7.6% and 105% for testosterone at 0.9 nmol/L and 0.9% and 107% for DHEAS at 0.2 nmol/L)
Excellent linearity: Calibration curves for the 9 steroid hormones showed r2 values above 0.99 across the calibration ranges for all the analytes
Introduction
Steroid hormones play vital roles in maintaining metabolic processes, immune regulation, and muscle integrity. Because these compounds exhibit extensive structural diversity and include many synthetic derivatives, precise determination of their concentrations in biological matrices is fundamental. Accurate analysis of steroid hormone levels enables a deeper understanding of hormonal balance and metabolic health, serving as a cornerstone for both clinical and biomedical research.
Methods
Sample preparation: Sample preparation was performed using Diagnotix’s steroid panel basic’ reagent set (https://www.diagnotix.com/nl/lcms-uhplc/Steroids) according to the manufacturer’s specifications. 100 μL calibrators in serum matrix were used to perform the SPE procedure. This reagent set is only available in certain EU countries.
Liquid chromatography conditions: Chromatographic separation was achieved using a Phenomenex Kinetex Biphenyl column (150 x 2.1 mm, 2.6 µm, 00F-4622-AN). Mobile phases A and B from the reagent set were used. The total run time was 10 minutes at a flow rate of 800 μL/min. The injection volume was 20 μL. The LC gradient program is presented in Table 1.
Results and discussion
Figure 1 shows the chromatographic separation of the panel of 9 steroid hormones at the lowest calibrator level. The 10 min gradient, in combination with the column selection and mobile phase composition, resulted in separation of the panel of steroid hormones. The extracted ion chromatograms (XICs) showed (S/N) values of 29:1 for corticosterone, 446:1 for cortisol, 55:1 for androstenedione, 35:1 for progesterone, 26:1 for 11-deoxycortisol, 16:1 for 21-deoxycortisol, 41:1 for 17-OH progesterone, 145:1 for testosterone and 1599:1 for DHEAS. at the lowest matrix calibrator measured, calculated using the peak-to-peak algorithm in SCIEX OS.
The quantitative performance of the method was investigated by injecting a series of calibrator samples spiked at the various concentration levels and injected in triplicate. Linearity, precision and accuracy were assessed across the calibration ranges for each of the 9 analytes. Figure 2 shows the representative XICs for A) corticosterone and B) 11- deoxycortisol across their respective concentration ranges (0.2-44.2 nmol/L for corticosterone and 0.4-45.7 nmol/L for 11- deoxycortisol). Figure 3 shows the calibration curves for vitamin B1 (left) and vitamin B6 (right) over the analytes’ respective calibration ranges. The plots show excellent linear responses across the calibration series, with r2 values greater than 0.99 for analytes.
The precision and accuracy values were calculated from 3 replicates of the lowest matrix calibrators analyzed. The precision (%CV) and % accuracy were 4.6% and 111% for corticosterone at 0.2 nmol/L, 3.9% and 104% for cortisol at 8.3 nmol/L, 2.4% and 108% for androstenedione 0.5 nmol/L, 0.7% and 110% for progesterone at 0.4 nmol/L, 8.0% and 107% for 11-deoxycortisol at 0.4 nmol/L, 1.2% and 92.0% for 21-deoxycortisol at 0.3 nmol/L, 7.4% and 98.6% for 17-OH progesterone at 0.3 nmol/L, 7.6% and 105% for testosterone at 0.9 nmol/L and 0.9% and 107% for DHEAS at 0.2 nmol/L.
Conclusion
A fast and accurate LC-MS/MS method for the detection of 9 steroid hormones extracted from human serum samples was developed using the SCIEX QTRAP 4500 system. The method demonstrated:
- Fast sample preparation, which consisted of a simple protein deproteination, requiring 100 μL of human serum
- Excellent sensitivity at the lowest calibrator level, resulting in signal-to-noise ratios (S/N) of 29:1 for corticosterone, 446:1 for cortisol, 55:1 for androstenedione, 35:1 for progesterone, 26:1 for 11- deoxycortisol, 16:1 for 21-deoxycortisol, 41:1 for 17-OH progesterone, 145:1 for testosterone and 1599:1 for DHEAS
- Excellent linear responses across the calibration series consisting of 6 calibrators, with r2 values greater than 0.99 for analytes
- High quantitation performance of the method, resulting in excellent precision from 0.7 to 8.0% and accuracy from 92.0 to 111% across the steroid hormones