Quantitative analysis of cortisol and cortisone in human urine using solid-phase extraction (SPE)
Abstract
This technical note demonstrates the accurate quantitation of cortisol and cortisone in human urine using a solid-phase extraction (SPE) procedure and analysis by the SCIEX QTRAP 4500 system. Low- to mid-nmol/L level sensitivity was achieved at the lowest calibrator with signal-to-noise ratios (S/N) of 1330:1 for cortisol at 34.0 nmol/L and 1688:1 for cortisone at 41.0 nmol/L. In addition, excellent linearity was observed across clinically relevant concentrations, with r2 values above 0.99 across the calibration range (34.0-601 nmol/L for cortisol and 41.0-693 nmol/L for cortisone).
Key benefits of cortisol and cortisone analysis from human urine using the QTRAP 4500 system
- Efficient sample preparation. Cortisol and cortisone were efficiently extracted from human urine samples using solid-phase extraction (SPE), starting with 100 μL of human urine
- Excellent linearity. Calibration curves for cortisol and cortisone showed r2 values above 0.99 across the calibration range (34.0-601 nmol/L for cortisol and 41.0-693 nmol/L for cortisone)
- Low-nmol/L level sensitivity and excellent quantitative performance. Sensitive quantitation for cortisol and cortisone was achieved with excellent precision (2.9% for cortisol and 0.4% for cortisone) and accuracy (101.7% for cortisol and 95.9% for cortisone) at the lowest calibrator levels (34.0 nmol/L for cortisol and 41.0 nmol/L for cortisone)
Introduction
Cortisol and cortisone play pivotal roles in governing the body’s stress response, metabolic pathways, and immune regulation. Their concentrations vary widely and can shift in response to both normal physiology and underlying disease, making reliable detection in human urine especially important. Robust quantification of these glucocorticoids offers valuable information about adrenal function and hormonal homeostasis, supporting a broad range of clinical assessments and research applications.
Methods
Sample preparation: Sample preparation was performed using Diagnotix’s cortisol & cortisone reagents set (https://www.diagnotix.com/en/products/cortisol-cortisoneurine-lc-ms-kit-2) according to the manufacturer’s specifications. 100 μL calibrators in human urine were used for the procedure. This reagent set is only available in certain EU countries.
Liquid chromatography conditions: Chromatographic separation was achieved using a Phenomenex Kinetex Luna Omega Polar C18 column (100 x 2.1 mm, 2.1 µm, 00D-4760- AN). Mobile phases A and B from the reagents set were used. The total run time was 5 minutes at a flow rate of 550 μL/min. The injection volume was 5 μL. The LC gradient program is presented in Table 1.
Results and discussion
Figure 1 shows the chromatographic separation of cortisol and cortisone in a control human urine sample at a final concentration of 34.0 nmol/L for cortisol and 41.0 nmol/L for cortisone, respectively. The extracted ion chromatograms showed a S/N of 1330:1 for cortisol and 1688:1 for cortisone, at the lowest matrix calibrator measured (34.0 nmol/L for cortisol and 41.0 nmol/L for cortisone), calculated using a peak-to-peak algorithm.
Figure 2 shows the representative extracted ion chromatograms (XICs) for A) cortisol and B) cortisone across their respective concentration ranges (34.0-601 nmol/L for cortisol and 41.0-693 nmol/L for cortisone). The peaks shown for the lowest matrix calibrators measured (34.0 nmol/L for cortisol and 41.0 nmol/L for cortisone) are well above the blank for both analytes.
The quantitative performance of the method was investigated by injecting a series of calibrator samples spiked at concentrations ranging from 34.0-601 nmol/L for cortisol and 41.0-693 nmol/L for cortisone, respectively. Linearity, accuracy and precision were assessed across the calibration ranges for each of the two analytes. Figure 3 shows the calibration curves for cortisol (top) and cortisone (bottom) over the analytes’ respective calibration ranges. The plots show excellent linear responses across the calibration series, with r2 values greater than 0.99 for both analytes.
The accuracy and precision values were calculated by 3 replicates in matrix at the lowest matrix calibrators measured (34.0 nmol/L for cortisol and 41.0 nmol/L for cortisone). The accuracy was 101.7% for cortisol and 95.9% for cortisone, respectively. The precision (%CV) was 2.9% for cortisol and 0.4% for cortisone.
Conclusion
A fast and accurate LC-MS/MS method for the quantitation of cortisol and cortisone extracted from human urine samples was developed using the SCIEX QTRAP 4500 system. The method demonstrated:
- Sample preparation using solid-phase extraction (SPE), starting with 100 μL of human urine
- Excellent sensitivity at the lowest calibrator level, resulting in S/N of 1330:1 for cortisol at 34.0 nmol/L and 1688:1 for cortisone at 41.0 nmol/L.
- Excellent linear responses across the calibration series consisting of 5 calibrators, with r2 values greater than 0.99 for both analytes
- High quantitative performance of the method, resulting in excellent precision (2.9% for cortisol and 0.4% for cortisone) and accuracy (101.7% for cortisol and 95.9% for cortisone) at the lowest calibrator levels (34.0 nmol/L for cortisol and 41.0 nmol/L for cortisone)