Option to disable the automatic calibration between different MS Methods
Due to the automatic calibration between different MS methods in the same batch (SWATH and IDA) an auto-calibration is performed after two samples instead of after every fourth sample as defined in the settings (our sample layout consists of one SWATH, two IDA and another SWATH acquisition for each sample). This leads to a highly increased consumption of the calibration solution and longer acquisition times. Manually removing the unnecessary auto-calibrations or only submitting the first of the four samples with the auto-calibration enabled is an option but very time consuming and error-prone for larger batches. On TripleTOF instruments the automatic calibration is strictly performed after the defined sample count (e.g. after every fourth sample regardless of the used MS methods). Therefore to match the auto-calibration sequence of TripleTOF instruments please add an option to disable the automatic calibration between different MS Methods.
0 Comments
Jul 29, 2019
33 Views
Product: SCIEX OS Software
Posted: now