Issue Description
The script - Manually Integrate - is missing in Analyst® software versions 1.7.2 and 1.7.3, although it was previously supported in versions 1.7 and 1.7.1. Analyst software has various peak-finding algorithms that identify and integrate peaks, but the Manually Integrate script can be useful to employ when a peak of interest is missed or only partially integrated. Scripts that are not automatically installed during the initial installation of Analyst software are usually found in C:\Program Files (x86)\Analyst\Scripts and can be installed manually. However, the Manually Integrate script installation folder is missing from this directory in version 1.7.2 and 1.7.3.
Resolution
Installing the Script
The installation files for the Manually Integrate script can be copied from the installer folder for Analyst® 1.7 or 1.7.1. (Access to a download of the older versions of Analyst software can be provided by technical support.) Prior to adding the script files, users must exit the Analyst software (1.7.2 or 1.7.3) and stop Analyst services.
Navigate to the Manually Integrate script installation files in the installer folder, Analyst 1.7.1 Software, located in the path: Analyst171-WebRelease\Scripts\Manually Integrate. The Manually Integrate folder contains an Install folder with three files - Files.Install, Manually Integrate.dll, and Manually Integrate.exe - and the ScriptRunner.exe file. The entire Manually Integrate folder must be copied to <Drive>:\Analyst Data\Projects\API Instrument\Processing Scripts. Open the Manually Integrate folder and double click on ScriptRunner.exe to install the script. Manually Integrate should now appear as an option in the script menu in Analyst software 1.7.2 or 1.7.3.
Using the Script
To use the Manually Integrate script, users must perform the following steps in Analyst software in explorer mode:
1. With a chromatogram opened, click Script > Manually Integrate. The Manual Integration 1 dialog is shown.
2. Select Options... to modify the fields in the Result section. In the Text File Exporting section, click Select to navigate to a text file.
3. Do one of the following: