Answer

The top-down and bottom-up proteomics workflows are well established for characterizing either intact proteins or peptide fragments after proteolytic digestion, respectively. However, the middle-down workflow for electron-activated dissociation (EAD)-fragmentation on a ZenoTOF® 7600 system is a new approach that efficiently characterizes relatively large protein fragments and IgG subunits by obtaining higher sequence coverage than traditional approaches that relied on multiple fragmentation techniques. EAD MS/MS experiments produce extensive, information-rich fragments that provide better spectral coverage for larger peptide and subunit sequences than collision-induced dissociation (CID) MS/MS. When coupled with the advanced algorithms found in Biologics Explorer software that can analyze, interpret, and annotate the more complex MS/MS EAD data, complete sequencing of larger peptides and subunits can be obtained in a single injection. 

An overview of the EAD-based middle-down workflow from sample preparation to data acquisition and analysis is shown below.
User-added image

In the EAD-based middle-down workflow (pictured above), an intact NIST mAb was proteolytically digested into the F(ab)’ and Fc/2 subunits; subsequent reduction produced the subunits Fc/2, LC and Fd subunits, which were chromatographically separated and then fragmented by EAD in a ZenoTOF 7600 system. Biologics Explorer software was used to interpret the EAD MS/MS spectra of the mAb subunits, producing an annotated map of the sequence.

The picture below shows the chromatograms of the subunit proteins (Fc/2, LC, and Fd) and the MRM HR-specific charge states of the protein envelope. Choosing a low, medium, and high charge state of the protein envelope is critical to get the most diverse fragmentation. The example below uses three different charge states to create an sMRM HR method.
User-added image
A typical sMRM HR acquisition method with EAD fragmentation is shown below for the chosen charge states (above). The red circled parameters are specific for a middle-down high resolution MRM acquisition method that employs EAD fragmentation and are a good starting point for a successful middle-down approach.

User-added image
User-added image
User-added image

Finally, the data files for each subunit can be processed in Biologics Explorer. All results from the middle-down analysis are loaded into the easy-to-use, optimized review template after processing in the workflow template. By reviewing the results in one template, they can be compared using a sequence coverage map, summary table and a combined MS/MS spectrum. The review template also facilitates sequence confirmation or PTM locations through the comparison of MS/MS spectra by overlaying or stacking the spectra using mirror plots.