Issue Description
Users of the TripleTOF® 6600 or 5600 systems and Analyst® TF1.7.1 software may experience intermittent acquisition failure, and a TDC error will appear in their application event log files: “Noise failing sync signal ignored" (ATF-665).
Resolution
There is a patch for the noise sync signal issue in Analyst® TF 1.7.1 software that addresses the “Noise failing sync signal ignored" warnings reported in the application log file.
The Analyst® TF 1.7.1 software has not undergone extensive formal testing at this time (2017), and this patch will be incorporated in the next version of the software after more rigorous testing. SCIEX assumes no liability or responsibility for any loss howsoever caused in connection with using this software. This patch is distributed upon request.