If a data file was acquired using simplified CAD gas settings and opened in the Analyst® software without the correct hardware profile active, the CAD gas setting in the file information will be displayed incorrectly.
A CAD gas setting of Low will be displayed as -1, Medium as -2, and High as -3. The CAD settings are displayed as they are stored in the software code (known issue, AN-299).