Adjust the regional language in LibraryView™ software to correct the SCIEX L:icense Notification screen


Date: 03/02/2026
Categories: SCIEX OS software

   Print    Rate Article:

For research use only. Not for use in diagnostic procedures.


Issue Description

When launching LibraryView™ software, the license notification screen appears, indicating that some library licenses were valid and other library licenses were not found. This screen appears primarily when two separate languages are used in SCIEX OS and LibraryView software simulataneously. The screenshot below shows an example of the error in Japanese:

Japanese lic notification.PNG


Resolution

To resolve this error, the same language must be selected in both SCIEX OS and LibraryView™ software.
The following languages are supported with SCIEX OS software 3.0 and higher: French, German, Italian, Japanese, Korean, Portuguese, Spanish, and simplified Chinese. 

To change the selected language in LibraryView software. open the software and select the cog on the upper right hand corner.
zoom -cog.PNG

Next, select Application Settings under the Regions tab. Choose the correct language (e.g., Japan).

lib _language settings.PNG

To select the regional language In SCIEX OS software, navigate to the General menu under the Configuration tile, and select the correct language from the drop-down menu.

SCIEX OS languag Settings.PNG

Choosing the correct language settings prevents to show the license notifications popup window.

For SCIEX OS systems set up for acquisition, only the English version of the operating system is supported. For processing versions of the software (SCIEX OS-Q and SCIEX OS-MQ), the English, French, German, and Italian versions of the operating systems are supported.