| Date: | 03/02/2026 |
| Categories: | Pharma CRO , DiscoveryQuant software |
For research use only. Not for use in diagnostic procedures.
Issue Description
Compound parameters could not be uploaded in Analyze mode in DiscoveryQuant™ 3.0.1 software after switching from a local database to SQL database due to issues with the database selected. The data had never been uploaded into the SQL database before; therefore, when the SQL database was chosen as the review database, no database could be referenced causing an error.
Resolution
To resolve this issue, manually upload data from the local database to the SQL database and then try uploading the compound parameters again. To alleviate the need for manual upload in the future, the file autoupload.txt was added to the folder C:\DiscoveryQuant30.
Comment