How to change company name in the MultiQuant software audit trail report ?


Date: 03/02/2026
Categories: MultiQuant software

0 Votes
   Print    Rate Article:

For research use only. Not for use in diagnostic procedures.


Answer

As shown in the picture below, the audit trail report associated with the Multiquant software with CFR features automatically indicates the company name.

image.png

Here are two procedures to modify the company name :

Procedure n°1 : Uninstall and reinstall Analyst® and MultiQuant software following the procedures provided by Sciex. Reinstalling Analyst® is necessary because the CFR functions of MultiQuant are linked to Analyst®. When reinstalling Analyst® software, enter the new company name in the window below when it is available in the Analyst installation process.

image.png

Procedure n°2

1) Go to Registry Editor of Windows. Administrator rights must be required to access the Windows registries and it is imperative to strictly respect the rest of the procedure to avoid making any unwanted modifications to Windows.

image.png

2) Go to Computer\HKEY_LOCAL_MACHINE\SOFTWARE\WOW6432Node\PE Sciex\Analyst\Install

3) Change the value of "AnalystRegisteredOrganization" with the new company name and close the registry editor.

image.png