How Can I Create a Custom Enzyme in BioPharmaView™ 3.x Software?


Date: 03/02/2026
Categories: Academia Omics , Biologics Explorer software

   Print    Rate Article:

For research use only. Not for use in diagnostic procedures.


Answer

This article describes how to use a different DataDictionary file to add a custom enzyme to BioPharmaView™ software. To make this change, the BioPharmaView.Data dictionary file cannot be directly modified. This will result in a reinstall of the software. Locate the file ProteinPilot.DataDictionary.xml in the directory C:\Program Files\AB SCIEX\ProteinPilot and make a copy. If ProteinPilot™ software is not installed, then request a trial license from our technical support team.
  • Open the ProteinPilot.DataDictionary.xml with Notepad or Notepad++.
  • Create the digestion set by searching for "DIGEST_SET" in Notepad. 
  • Copy an existing digest set and insert it underneath the original digest set.
  • Ensure the DIGEST_SET : XX (number is unique). Type in the enzyme.
  • Search the document to ensure the three-letter code “TLC” for this enzyme is unique.
  • Enter the amino acids and their unique cleavage attributes.
User-added image
  • The attribute table will tell you if you need to use CA or CB or CNB:
User-added image
  • Save this file and copy the ProteinPilot.DataDictionary file to C:\Program Files\AB SCIEX\BioPharmaView location. 
  • In addition, there will be a BioPharmaView.Data Dictionary file. This file cannot be modified. Please do not attempt to modify it otherwise the BPV software will need to be reinstalled.
  • To confirm that the newly added enzyme can be viewed in BPV software, open any project and choose a peptide setting workflow.
User-added image