How To Output Qualifier Transition Data Using Reporter 3.2 Software


Date: 03/02/2026
Categories: SCIEX OS software , Analyst software , MultiQuant software

   Print    Rate Article:

For research use only. Not for use in diagnostic procedures.


Answer

Using Reporter 3.2 software is a great way to report out specific types of data selectively, including transition-specific data for qualifier ions. To output specific data for qualifier ions, the first option is to use an IF tag around the information that should be for qualifying transitions. In this example, the mass ranges tag is used. The IF tag will need the syntax below for the condition:

EXACT($Component_Type,'Qualifiers')
solo qual IF tag.PNG
The same IF tag can be used around multiple tags within a loop to selectively qualify multiple tags:
qual IF tag.PNG

Another option is to use the ratio ions loop.
qual ratio ions.PNG

Both options require the quantifier and qualifier transitions to be grouped in the processing method. The Reporter 3.2 software assumes that the component names ending in 1 are the quantifier transitions, and those ending in 2 are the qualifier transitions (e.g., PFOS-1 and PFOS-2).