How to Report Regression Parameters in a Results Table Created in SCIEX OS Software


Date: 03/02/2026
Categories: SCIEX OS software

   Print    Rate Article:

For research use only. Not for use in diagnostic procedures.


Answer

Regression parameters values (e.g., r squared, slope, linear coefficient) can now be added to a results table created in SCIEX OS software from the calculated columns that are saved under the processing method. 

For example, to add the r squared value to a results table, open the processing method, navigate to the Calculated Columns section and then select Add Formula. In the field, type in the Formula Name, and then under Formula Details choose the regression parameter. In this case for the r squared value, choose [r^2]. Other regression parameters can be added as a custom column to the results table in the same way.

r squared.PNG