How To Provide Selective Access to Analyst® Software Projects


Date: 06/11/2020
Categories: Analyst software

   Print    Rate Article:

To set access for select users at the project level in Analyst® software, the security configuration must be updated by following these steps:

  1. Open the Security Configuration and ensure that the Security Mode option is set to Mixed Mode
  2. On the People tab, enter all users and their roles
  3. On the Projects Tab, highlight the project folder of interest and then select View/Edit Access Rights
  4. Click Edit and then Add and then select the users who will have access to the project
  5. On the Permissions for Everyone section below, add the permissions granted for each user
  6. For any other users in the Analyst database that should not have access to the project, select Deny on the Permissions screen
  7. Repeat as needed for additional users and projects

When trying to access another person’s projects, users without the correct access will get the following message:"Access is denied."