Setting up a Rack Changer in Sciex OS Simulation Mode


Date: 03/02/2026
Categories: SCIEX OS software

   Print    Rate Article:

For research use only. Not for use in diagnostic procedures.


Answer

To set up the rack changer in Sciex OS in simulation mode:

  1. Open Sciex OS software, and then select Devices under the Configuration mode. 
  2. Click on Add, and then choose Integrated System as the Type and Sciex Shimadzu RUO as the Model.
  3. Click Settings, and then select check Simulate Device. 
  4. Click the the grey plus sign to add the Autosampler type. 
  5. Select Installed Rack Changer. 
User-added image