For research use only. Not for use in diagnostic procedures.
Answer
When you create a reprocessing method in SCIEX OS you can enter a list of targeted compounds by entering different information such as the mass of the precursor and the fragment, the retention time, etc.
When you do not know in advance the retention time of a compound, you can use the "Find X peaks" feature by clicking on the drop-down menu in the Retention time mode column.
In this drop-down menu:
When using the "Find X peaks" feature, the software may not select the most intense peak. This may be due to a calibration issue with the mass spectrometer as explained in the example below:
In the chromatogram above Sciex OS does not select the most intense peak at 1.75 min which corresponds to morphine. If we manually select the peak which corresponds to morphine, a mass error of 41 ppm is indicated on the precursor (0.012 Da). while the XIC width configured in the Components table for Morphine was 0.02 Da (±0.01 Da).
Due to the elevated mass error, the Morphine XIC fell outside the defined extraction window, resulting in the peak at RT 1.75 min being excluded during the “Find 5 Peaks” algorithm execution.
As an immediate workaround, we recommend increasing the XIC Width setting for Morphine in the processing method’s Components page from 0.02 Da to 0.04 Da. This adjustment restores the expected peak detection behavior, as the observed mass error of 0.012 Da will fall within the expanded window.
For long-term stability and optimal performance, we also recommend performing a full instrument recalibration.