Answer
The ACMP Peltier effectivity is based on the ambient temperature of the instrument. The operational range of the ACMP is between 4-40C. In order to reach 4C, the room temperature needs to be 30C or lower with a relative humidity of less than 70%. However, the temperature accuracy is +/-6C.