Incorrect Argument Error when Creating MultiQuant™ Software Results Table


Date: 03/02/2026
Categories: MultiQuant software

   Print    Rate Article:

For research use only. Not for use in diagnostic procedures.


Issue Description

When trying to create a results table in MultiQuant™ software, the following error message appears: "Incorrect Argument: Length cannot be less than zero. Parameter name: length."

MultiQuant incorrect argument

 


Resolution

This error can be caused when individual .wiff and .wiff.scan files from Analyst® software are cut and pasted directly onto a thumb drive (or another data transport device) outside of the project folder structure. The data should be transferred from one computer to another by moving the entire project rather than individual data files.

Try transferring over the entire project from Analyst to the processing workstation where MultiQuant is installed and then creating the results table using data files stored in the project folder.