Date: | 10/23/2017 |
Categories: | Mass spectrometry , Analyst software |
For research use only. Not for use in diagnostic procedures.
Issue Description
If the Regional Settings on an Acquisition Computer with Analyst® Software is not set to US English, this can cause the Error message: 'Batch not submitted-Injection volume must be a positive value' when you try to submit a Batch.
Resolution
Please change the Regional Settings of your Computer. Could be done over the Windows Start Icon and choose "Control Pane", than "Regional Settings" or similar in your own Languages. If you don't like to change the whole Settings to US English, you will need to change the Decimal Separators to a period (.). Be aware, that this change will also influence other Software like Microsoft Excel.
Posted: now