LibraryView 1.0 Error 21467259 failed to create SQL database Unknown Error


   Print    Rate Article:

For research use only. Not for use in diagnostic procedures.

Issue Description 

When installing LibraryView 1.0 the following error message is shown. Error -21467259: failed to create SQL database: LibraryView, error detail: Unknown Error.

Resolution 

This is because LibraryView 1.0 is not supported on a 64bit Operating System.  The solution is to install on 32bit operating system only.

From LibraryView Release Notes

 

Requirements

• Windows XP operating system with SP3 or the Windows 7 (32-bit) operating system, with or without SP1.

• Microsoft Office 2007 or 2010.

• To create .pdf versions of reports when Microsoft Office 2007 is installed, the 2007 Microsoft Add-in: Microsoft Save as PDF must be installed. The free download is available from www.microsoft.com/download.

• To create acquisition and quantitation methods using the LibraryView software, Analyst® software version 1.6 or higher must be installed.

RUO-MKT-18-2713