Issue Description
To avoid Problems with 32 Karat Software and its associated errors in transport failures the PA800, PA800
Enhanced and the PA800 PLUS systems have the possibility to perform a Master- Reset. This also applies to MDQ and MDQ PLus
This reset has no influence on Software/System Qualification Status and can be done when needed
and/or on a regular basis (e.g. weekly). With the reset you are not deleting any analysis/system/user
data.
Resolution
1. Switch off device first
2. Press button on back-side of device with a plastic part (e.g. Pipette tip). To avoid short-circuiting
do not use any metal or conductive part (e.g. Pen, screw-driver).
3. Hold the button on back side together with the power button in front pressed for at least 5-6 sec. A beep sound will occur which is normal, which will end as soon as you release both buttons
4. Now let the instrument initialize