MultiQuant™ Software Report Error: Object Reference Not Set to an Instance of an Object


Date: 03/02/2026
Categories: MultiQuant software

0 Votes
   Print    Rate Article:

For research use only. Not for use in diagnostic procedures.


Issue Description

When creating a report in MultiQuant™ software, the following error appears: "Object reference not set to an instance of an object."


Resolution

This error results because a file other than a report template was selected in the Create Report dialog box. Check that the file being used in the Report template field is a true report template. Open the file to see if there are any tags. Most likely, the file chosen to be the template is a generated report and not a report template. Navigate to the MultiQuant software report template folder (C:/ProgramData/AB SCIEX/MultiQuant/Reporter) and select a report template with tags to create a report.