| Date: | 03/16/2026 |
| Categories: | Cliquid software |
For research use only. Not for use in diagnostic procedures.
Issue Description
After installing Cliquid software version 3.2, you are unable to create or select report styles. This is caused by a missing folder that is not created during installation which is needed to create a report style in Cliquid software and generate reports.
Resolution
Create a folder named Report Templates in the following folder: <Analyst Data>\Projects\Cliquid\Templates.
Once the Report Templates folder is created, place all docx templates into this folder. You will now be able to set report template styles in Cliquid software.
Comment