How to add custom columns to a Report Template in Reporter 3.2 software


Date: 03/12/2026
Categories: Analyst software

   Print    Rate Article:

For research use only. Not for use in diagnostic procedures.


Issue Description

After custom columns were added to an acquisition batch file, those columns with their appropriate data could be seen in the Analyst software results table.  These custom columns need to be added to the report template as well. Adding a text tag to the report template with the type set to Custom did not allow the custom data to be shown in the report.


Resolution

To add a value from a custom column, one needs to add TextField with “Custom” field, as seen below.
For Reporter software, custom column names cannot have spaces.

Example: A custom column, Instrument Name, needs to be renamed to "Instrument_Name" or "InstrumentName" to make it work in Analyst Reporter 3.2 when reports are generated through Analyst software.

Note: if MultiQuant software is used to generate reports, spaces are allowed in custom column names.

User-added image