SCIEX OS Software: Internal Standard Data in Report Does Not Match the Results Table


Date: 03/02/2026
Categories: SCIEX OS software

   Print    Rate Article:

For research use only. Not for use in diagnostic procedures.


Issue Description

If a query is used along with a report template to create a report in SCIEX OS software 1.6.1 or lower versions, an issue has been observed for reports that include internal standard samples--where there is a mismatch with the index in the Excel query file. This results in the custom tags in the report template incorrectly reporting values or not reporting any values generated in by the query.
 


Resolution

To resolve this issue, all internal standard samples must be listed at the bottom of the SCIEX OS software processing method in the Analytics mode. To ensure this, open the processing method and select the components tab. The internal standards should only be selected at the bottom of the components list and not at the top or embedded between the analytes in the processing method.