Ensuring the Positive Hits Report Includes Only Positive Hits from a Results Table Generated in SCIEX OS Software


Date: 03/02/2026
Categories: SCIEX OS software

   Print    Rate Article:

For research use only. Not for use in diagnostic procedures.


Issue Description

The Positive Hits report template is not selectively reporting the positive hits in the Sciex OS software results table that was generated using a non-targeted or targeted library, and the template is instead reporting out all generated data.


Resolution

A report generated using the Positive Hits template will include data from all visible rows on a results table and will not automatically select for just the positive hits to appear in the report. In order to show only data corresponding to positive hits on a report, the results table itself needs to be filtered first for the positives. Once the results table is filtered, then a report showing only positive hits can be generated using the Positive Hits report template.