Answer
The six-peptide mixture (part number 4465940) is used for the calibration of mass spectrometers. The kit contains the following six peptides and their mono-isotopic masses: bradykinin (2-9 clip) (904.4681), angiotensin I (1296.6853), glu1-fibrinopeptide (1570.6774), ACTH (1-17 clip) (2093.0867), ACTH (18-39 clip) (2465.1989), and ACTH (7-38 clip) (3657.9294).
Here are the sequences of the peptides found in the six-peptide mixture:
Peptides Sequence
Bradykinin 2-9 PPGFSPFR
Angiotensin I DRVYIHPFHL
Glufib EGVNDNEEGFFSAR
ACTH (1–17 clip) SYSMEHFRWGKPVGKKR
ACTH (18–39 clip) RPVKVYPNGAEDESAEAFPLEF
ACTH (7-38) FRWGKPVGKKRRPVKVYPNGAEDESAEAFPLE