What Fittings Are Needed To Add a Waters M-Class LC to My ZenoTOF 7600 System?


Date: 03/02/2026
Categories: QTOF systems

   Print    Rate Article:

For research use only. Not for use in diagnostic procedures.


Answer

PEEKSil tubing is normally used to connect a Waters ACQUITY M-Class UPLC to a ZenoTOF 7600 system. The PEEKSil tubing has an outer ID of 1/32 inch, but the tubing for the Waters M-Class has an outer ID of 1/16 inch, so purchasing the correct fittings are critical in order to connect LC components, such as trap columns, to the ZenoTOF 7600 system.

The recommended hardware for making the connection between the ZenoTOF 7600 system and the M-Class UPLC tubing are:

  • SCIEX P/N 50198215: 5 PK, Fitting,1/32 in PEEKSil to 1/16 in 10-32 Port
  • SCIEX P/N 5019820: 5 PK, Ferrule,1/32 in PEEKSil to 1/16 in 10-32 Port