Why Do the Calculated Concentration Values Show as Degenerate in SCIEX OS Software?


Date: 03/02/2026
Categories: SCIEX OS software , MultiQuant software

   Print    Rate Article:

For research use only. Not for use in diagnostic procedures.


Answer

In SCIEX OS software, if a value in the Calculated Concentration column displays as degenerate, this is because the values for Actual Concentration for standard samples have not been added to the results table.

User-added image

Type in values in the Actual Concentration column for each standard, and then the calculated concentration for each standard will be calculated. In order to have the actual concentration values appear automatically in the results table,  this information must be added to the batch prior to acquiring the data.

User-added image