Why Do I Receive Memory Concerns Warning when Processing Data in MultiQuant™ Software?


Date: 03/02/2026
Categories:

   Print    Rate Article:

For research use only. Not for use in diagnostic procedures.


Answer

When processing a large data set in MultiQuant™ software 3.0.3, users may see a Memory Concerns message, "The Results Table you opened may cause out of memory computer issues. To free-up memory close other applications and any other opened results table."

User-added image

This message is a warning, not an error message, regarding the amount of memory required to process and display a large data set. This error message will be generated for large data sets for both GLP and non-GPL versions of MultiQuant software, but the GLP version will throw the error for results tables with lower numbers of rows (40.000) compared to the non-GLP results tables (100,000 rows).