Abstract
Kinetic reactions are critical for understanding how compounds behave in changing physical and chemical environments. Traditionally, analyzing samples from kinetic studies might involve quenching, aliquoting and transporting to instruments through the laboratory. Here, we utilized the Beckman Coulter® Life Sciences Biomek i7 automated liquid handler to prepare and perform a kinetic study and tasked the Echo® MS+ system with ZenoTOF 7600 system to rapidly quantify the analytes.
Key features of the automated real-time kinetic hydrolysis of codeine-6-glucuronide
- Automated sample preparation: Pipette, incubate and shake the kinetic study samples using the Biomek i7 automated liquid handler
- Real-time Zeno MRMHR analysis: Analyze the study samples in wide peak mode with Zeno MRMHR at 5 seconds per sample
- Quantitative results: Quantify targeted analytes and internal standards at specific time intervals
- Evade quenching: Rapidly analyze study samples without modifying the chemical environment
- Avoid sample cleanup: Analyze samples directly, even in the presence of a hydrolysis enzyme
Introduction
Kinetic assays are time-consuming and require accurately timed sample preparation. Furthermore, lengthy analysis methods such as liquid chromatography can impact the results of kinetic studies.
Codeine is a naturally occurring opiate that can be consumed orally to treat acute pain via the conversion of codeine to morphine in the liver.1 The codeine that is not converted to morphine undergoes glucuronidation or N-demethylation.1 Codeine-6-glucuronide, a product of codeine glucuronidation, is difficult to convert to codeine, even using beta-glucuronidase enzymes.2,3 In this study, we used the Biomek i7 automated liquid handler from Beckman Coulter Life Sciences and the Echo® MS+ system with ZenoTOF 7600 system to monitor the conversion of codeine-6-glucuronide to codeine. The conversion of a 50 ng/mL and of a 250 ng/mL sample of codeine-6- glucuronide was monitored at 5 time points over 1 hour while incubating at 55°C (Figure 1).
Sample preparation was programmed on the Biomek i7 automated liquid handler for a kinetic study. Pause steps were built into the method so that the Echo®-qualified 384-well plate could be removed from the Biomek i7 platform and transferred directly to the Echo® MS+ system with ZenoTOF 7600 system for rapid analysis at specified time points.
Methods
Automated sample preparation: The Beckman Coulter Life Sciences Biomek i7 automated liquid handler performed the pipetting, shaking and incubation steps. The reaction was carried out in a 96-well 1 mL plate throughout the kinetic study (Figure 2). The device transferred 50 µL of reaction mixture from the 96- well plate to a 384-well plate and was then paused at 0, 10, 20, 30 and 60 minutes so that the user could remove the 384-well plate from the Biomek i7 automated liquid handler and transfer it to the Echo® MS+ system with ZenoTOF 7600 system for a rapid, real-time analysis at each time point.
When the Biomek i7 automated liquid handler was paused for plate removal, the first datapoint (0 minutes) was collected on the Echo® MS+ system with ZenoTOF 7600 system. Following data acquisition, the Echo®-qualified 384-well plate was returned to the Biomek i7 automated liquid handler and the sample preparation program was resumed. The Biomek i7 automated liquid handler then transferred the 96-well reaction plate to an integrated BioShake Q1 (QINSTRUMENTS, Jena, Germany) on board the Biomek i7 automated liquid handler deck. Here, the 96-well reaction plate was shaken at 850 RPM for 10 minutes while incubating at 55°C. After shaking and incubating, the 96-well plate was transferred to an open position on the Biomek i7 automated liquid handler deck. A 50 µL aliquot of each test and control sample was taken from the 96-well plate and transferred to the 384-well plate. The Biomek i7 automated liquid handler was paused and the 384-well plate was removed for sample analysis on the Echo® MS+ system with ZenoTOF 7600 system (10-minute datapoint). The 384-well plate was then returned to the Biomek i7 automated liquid handler and the above process was repeated to collect the remaining datapoints at 20, 30 and 60 minutes.
Acoustic ejection: The carrier solvent was 80:20 (v/v), acetonitrile/water with 0.1% formic acid and was introduced at a flow rate of 460 µL/min. A total of 70 nL of each control and test sample was ejected in wide peak mode for 5 seconds at 10 Hz.
Mass spectrometry: A Zeno MRMHR method was utilized to analyze codeine, codeine-d6 and codeine-6-glucuronide. Precursor ions with m/z 300.1597, 306.1976 and 476.1912 were analyzed for codeine, codeine-d6, and codeine-6-glucuronide, respectively. A TOF MS/MS scan was performed for product ions of the precursor ions between 100–500 m/z. Two product ions were chosen to quantitatively measure each of the above ions (Tables 1-4).
Data processing: Data processing was performed using the Explorer and Analytics modules of SCIEX OS software.
Method optimization
The Zeno MRMHR scan generated product ions of the chosen precursor ions. The product ions used for quantitation in the Analytics module of SCIEX OS software were selected from the data generated in the Zeno MRMHR method shown in Figure 3.
Quantitative kinetic study
The calibration curves for the primary and secondary codeine transitions were linear from 10–250 ng/mL with r >0.998 (Figure 4).
Conclusion
- A 1-hour kinetic study was conducted with pipetting, incubation and agitation performed in a reaction plate on the Biomek i7 automated liquid handler
- A Zeno MRMHR analysis of free codeine was performed at 10- minute intervals for high quantitative sensitivity
- Samples were analyzed at a rate of 5 seconds per sample without disrupting the timing of the kinetic study
- Free codeine increased incrementally in the 50 ng/mL and 250 ng/mL samples as codeine-6-glucuronide was enzymatically converted to codeine within 1 hour
- No quenching was required
- No sample cleanup was required, even with enzyme present in the wells
References