Native State Charge Variant Analysis of Commercialized Monoclonal Antibodies in Minutes


CZE Rapid Charge Variant Analysis Kit

Peter Holper, SCIEX
Redwood City, CA 

Abstract


Monitoring of variants for therapeutic proteins and antibodies is required throughout manufacturing to assess drug purity and heterogeneity and to ensure the safety and efficacy of the drug.  Of all the analytical methods to be developed, charge variant analysis methods used to monitor post-translational modifications (PTMs), require the longest time commitment.  Presented in this technical note is how the SCIEX CZE Rapid Charge Variant Analysis Kit (P/N C44790) can be used for charge variant analysis using Capillary Zone Electrophoresis (CZE) which combines the benefits of a native state analysis with the speed and resolution expected of a capillary electrophoresis method.

RUO-MKT-02-10446-a
Figure 1. (A) The PA 800 Plus Pharmaceutical Analysis System, (B) the CZE Rapid Charge Variant Analysis Kit (P/N C44790) and (C) pre-assembled cartridge (P/N A55625).

Introduction


As biopharmaceutical companies face increasing pressure to decrease the overall drug development timeline, analytical researchers find themselves finalizing methods earlier in development with less material on hand. Of all the analytical methods to be developed, charge variant analysis methods used to monitor post-translational modifications (PTMs), require the longest time commitment. Quickly finding a robust, stabilityindicating method that can monitor charge variants is an ongoing analytical challenge.

HPLC-based ion-exchange chromatography (IEX) methods commonly analyze biotherapeutics in their native state. These methods can be optimized with regard to column stationary phase, mobile phase pH, type of and concentration of buffer salts, gradient conditions, and a host of other parameters. Development of a suitable method for a single molecule can easily span weeks to months.

Capillary Isoelectric Focusing (cIEF) charge variant methods generally analyze biotherapeutics in a denatured state. Development time is often much shorter compared to IEX due to fewer parameters to optimize, including buffer additives, concentration and focusing time. However, when analyzed in a denatured state, generated stability data does not always trend as well as native state stability data.

Capillary Zone Electrophoresis (CZE) combines the benefits of a native state analysis with the speed and resolution expected of a capillary electrophoresis method. Using a widely universal buffer,1 this technique requires little to no method development, has a simple sample preparation and a higher analytical throughput than IEX or icIEF.

The SCIEX CZE Rapid Charge Variant Analysis Kit (P/N C44790) provides high resolution, high sensitivity CZE separations.

Charge Heterogeneity in Minutes


• Quantify charge variants in their native state

• Little-to-no method development

• Dilute, shoot and prepare for your next CE application

• Platform capable methodology

Figure 2. Fullview and zoomed electropherogram of Herceptin® (trastuzumab).

Figure 3. Fullview and zoomed electropherogram of Rituxan® (rituximab).

Figure 4. Fullview and zoomed electropherogram of Remicade® (infliximab).

Demonstrated here are the charge variant analyses of multiple commercially available monoclonal antibodies on the PA 800 Plus platform using this kit. This method can easily be applied to therapeutic proteins such as antibody-drug conjugates (ADCs), pre-clinical monoclonal antibodies (mAbs), and fusion proteins with little to no method development. Additionally, it can be easily modified to suit more specific user criteria. 

Methods


Instrumentation

All experiments were performed on the PA 800 Plus Pharmaceutical Analysis System (SCIEX). A Pre-Assembled Cartridge (SCIEX P/N A55625) was used for the separation of all three commercially available monoclonal antibodies: Herceptin® (trastuzumab), Rituxan® (rituximab) and Remicade® (infliximab).

Reagents

Reagents used to prepare the samples and analysis buffers were provided in the CZE Rapid Charge Variant Analysis Kit (SCIEX P/N C44790). The therapeutic proteins of Herceptin® , Remicade® and Rituxan® were purchased from Myoderm (Norristown, PA).

Sample Preparation

Herceptin® , Remicade® and Rituxan® were diluted to 1 mg/mL in CE Grade water (SCIEX P/N C48034).

Instrumentation

The capillary electrophoresis instrument used was a PA 800 Plus equipped with UV detection and a 214nm bandpass filter (SCIEX P/N 144437).

Separation and Analysis

Separations were performed at 1000 V/cm and injection was performed by pressure for 10 seconds at 0.5 PSI. The separation buffer used was from the CZE Rapid Charge Variant Analysis Kit (SCIEX P/N C44790). Data acquisition and analysis was done using 32 Karat software V10. 

Results


To investigate the reproducibility of the assay, 10 injections of each molecule were made. The peak profile was stable and %RSDs of the total relative area perentages of acidic variants, basic variants and main peak were below 2% (tables 1-3).

Table 1. Statistics of Percent Total Acidic variants, Percent Total Main Peak, and Percent Total Basic Variants of Herceptin® (trastuzumab).

Table 2. Statistics of Percent Total Acidic variants, Percent Total Main Peak, and Percent Total Basic Variants of Remicade® (infliximab).

Table 3. Statistics of Percent Total Acidic variants, Percent Total Main Peak, and Percent Total Basic Variants of Rituxan® (rituximab).

Conclusion


• Reproducible:
CZE separations are shown to be reproducible through both replicate injections of commercial mAbs and an inter-company collaboration.2

• Platform capable: The simplicity of separation and optimal pH of the buffer allows for application to molecules across a broad pI range above 7.0.2

• Limited method development: Additional resolution can easily be obtained through the use of a longer bare-fused silica capillary, lower capillary temperature or separation voltage. Further development has focused on buffer pH and levels of additives.3

• High-Throughput: High-resolution separation can be achieved in approximately 5 minutes or less. With only two minutes of buffer replenishment in between injections, total sample analysis time can be completed in under 10 minutes.

References


1. He, Y., Isele, C., Hou, W., Ruesch, M., J. Sep. Sci. 2011, 34, 548–555.

2. Moritz, B., Schnaible, V., Kiessig, S., Heyne, A., Wild, M., Finkler, C., Christians, S., Mueller, K., Zhang, L., Furuya, K., Hassel, M., Hamm, M., Rustandi, R., He, Y., Solano, O. S., Whitmore, C., Park, S. A., Hansen, D., Santos, M., Lies, M., J. Chromatogr. B 2015, 983–984, 101–110.

3. Moritz, B., Locatelli, V., Niess, M., Bathke, A., Kiessig, S., Entler, B., Finkler, C., Wegele, H., Stracke, J., Electrophoresis 2017, 38, 3136-3146.