LipidView software enables fast lipid characterization and quantification, using an easy 4-step workflow with processing and review of MS and MS/MS data from infusion and LC ESI experiments. An internal Lipid fragment database with more than 40 lipid classes and 23 000 lipid species supports identification of lipids species. Works with SCIEX Triple Quad system, QTRAP system, and TripleTOF system mass spectrometers.
Lipid profiling is performed using electrospray mass spectrometry, where the acquisition of MS and MS/MS data enables structural characterization of lipid species. These qualitative strategies require lipid fragmentation and spectral interpretation to identify lipid molecular species by head group, fatty acid, and carbon chain compositions. Lipid profiling can also provide quantitative information when a proper internal standard strategy is implemented. Through the use of complementary scanning techniques, such as multiplexed precursor ion scanning and data dependent acquisitions, lipids can be identified and their detection corrected by internal standards to allow for direct comparisons between many samples.
The processing of such complex and complementary datasets is predominantly the bottleneck in lipid profiling experiments of biological lipid extracts and thus only software consisting of sophisticated bioinformatic tools can allow for these analyses in high throughput. LipidView software provides even novice lipid investigators the ability to easily find, identify and interpret lipid species from mass spectrometry data from complex samples.
LipidView software provides a user-friendly workflow driven format for the identification and quantification of lipid species from ESI-MS data (Figure 1). The software enables identification and characterization through database searching of MS/MS fragment ions and reports a numerical and graphical output for profiling various lipid classes by lipid head groups, fatty acid and long chain base characteristic fragments and neutral losses. LipidView software streamlines the automated data processing with method selection, lipid species identification, isotopic correction of peak intensities, multiple internal standards-based quantification, lipid profile visualization, and reporting of the results.
Lipid catalogue utility
The Lipid Catalogue Utility provides access to the entire database of lipid classification covered by over 23 000 lipid species, retrieving molecular formula, monoisotopic mass, m/z of common adducts, fatty acids, and characteristic fragments listed in either positive or negative polarity. As shown in Figure 2A, the Lipid Catalogue Utility demonstrates the results retrieved from the input of PE 38:4, following the nomenclature: <headgroup> <sum of carbon atoms>:<sum of double bonds> to retrieve the database information for the desired lipid species, including the isotope correction factor implemented when processing data.
Lipid calculator utility
An alternative to retrieving lipid information by lipid name, the user can search lipid species by a given m/z value in either positive or negative polarity with a designated mass tolerance. Figure 2B shows the output of querying potential lipids by parent mass in positive mode listing all lipid species whose monoisotopic mass matches within 0.3 amu. The retrieved list can also be filtered to include only entries containing common fatty acid chain compositions.
The use of lipid class-specific internal standards is critical to making quantitative measurements or generating normalized comparisons between samples analyzed by any mass spectrometry experiment. LipidView software offers a module dedicated to the implementation and management of internal standards, as shown in Figure 3. The application contains a pre-loaded list of non-natural, internal standards (consisting of odd chained fatty acid moieties) commonly used as standards for glycerophospholipids, glycerolipids, sphingolipids, and sterol lipid classes consisting of odd chained fatty acid moieties. In addition, users can also enter custom internal standards by filling in the required information, such as molecular formula, molecular ion and specific scan information (Figure 3, inset)
Once internal standards have been added to the lipid database, users can then select specific internal standards within a processing method to apply when processing lipid mass spectrometry data.
LipidView software enables batch processing of ESI lipid data by applying processing methods to samples individually, or to a group of samples. The processing method set-up module is shown in Figure 4. Here a user defines the parameters for extracting the raw data and spectral peak pre-processing, then selects the relevant lipid classes for searching the data against the database. Processing methods are set-up by 5 primary steps, as outlined in the Figure 4 below.
Processing methods can be saved and are added to the list for future use. A set of pre-installed processing methods is available in the software; these methods serve as templates to derive user specific methods.
In the Spectrum Tab of the process method set-up module, the user can select whether the data should be smoothed (recommended) and isotopic contribution from lower mass species be removed before lipid identification is carried out. This feature is essential when processing data from complex samples where lipid species are not resolved in time to ensure accurate lipid identification from isobaric peaks especially lipids differing by one double bond.
Internal standard reference peaks can be used for correcting the peak areas or peak intensities of identified lipid species. Default internal standards are assigned to lipids based on the Default Lipid Class specified on internal standard information entry. To customize how the internal standards are to be applied, the user can open Profile Tests to view the complete list of identified lipids and review the automatic assignment of each internal standard to the applicable lipid species (Profile Tests Editor shown in Figure 5). Once all lipids have the appropriate internal standard assigned, the numerical data is normalized and quantitative lipid profiles can be created to compare lipid abundances between samples.
Once a data file has been processed by LipidView software and internal standards have been designated to correct a particular lipid-class or specific lipid species, the user can create a Target Lipid List consisting of a set of specific lipid targets and internal standard reference peaks to apply against the processing of new datasets. Target Lipid Lists can be saved and then loaded into the application when the user develops a processing method, as shown in upper view of the Processing Method dialog shown in Figure 4. Processing datasets using a Target Lipid List relies purely on the information pertaining to the targeted lipids and provides a defined search within the lipid fragments database. Such an approach is advantageous for the batch processing of multiple samples in lipid class-specific screening experiments, since it is fast, data can be aligned easily and species which are not described in the lipid database can be a part of the target lipid list.
Data files acquired from SCIEX Triple Quad system, QTRAP system or TripleTOF system mass spectrometers can be processed by LipidView software to identify lipid species through extraction of spectral peaks acquired via specific survey or MS/MS scans. Spectra from survey scans are processed independently from the dependent MS/MS scans (MS triggered MS/MS). However, survey scan and MS/MS lipid identification results are coupled within the results workspace. In addition, a user can choose to store information on spectral peaks that could not be assigned to a lipid species; such results are available in parallel as well.
Survey scan and MS/MS data review
Precursor Ion, Neutral Loss, Q1, Enhanced MS, MRM survey, TOF MS, IDA, or MSMSALL scans undergo peak extraction and direct searching of precursor masses against the lipid fragments database to obtain lipid identifications and spectral data shown in the results workspace (Figure 6A). Identified lipids can be presented in three categories: All, Confirmed, and Confirmed & Common.
Numerical data is aligned per lipid identification in the form of peak intensities, peak areas, % intensities, or representative values corrected by internal standards. MS/MS data collected by information dependent acquisition (IDA) is processed separately and is coupled to the results workspace yet viewed independently (Figure 6B). As MS/MS data is more likely to contain several confirmatory ions, the lipids identified by positive and negative MS/MS spectra are often well validated and can be reviewed and exported separately. Samples can be filtered and grouped where lipid profile plots are created using averaged data between sample groups. Moreover, there is a direct link between the sorted numerical data and the lipid profiles quick view within the results workspace.
Advanced lipid profiles
Lipid profiles can also be created in the Advanced Profiles View enabling the plotting of large numbers of lipids by species, by class, or by fatty acids in an expanded view. As shown in Figure 6C, the average corrected peak intensities are plotted for two sample groups for direct comparisons. Advanced Profiles View also allows the user to create secondary plots. For example, in Figure 6C, the total PC lipid class is plotted in the major view, but a second plot can be selected to show fatty acid composition of a selected fatty acid chain, highlighted by the red box. The data plotted in each profile is displayed below each graph for easy export of such plots.
Spectral view
To view the raw spectra, a Spectral View module allows for the viewing of specific scans or averaged spectra from processed data files. Figure 6D is a spectrum corresponding to the PIS of 283.3 m/z in negative mode for a liver tissue extract displayed subtracted from an identical spectrum from red blood cell (RBC) lipid extract. Spectral peaks loaded into the spectral viewer can also labeled with the corresponding lipids identified.
Corrected peak intensity, peak area data displayed in the upper grid of the results workspace, Advanced Profiles graphical values, or the lipids identified from the Profile Tests can be exported to a text file (*.txt). Exporting of data maintains the sample grouping or filters created within the LipidView software Session. Also, samples processed maintain all experimental details, such as parent mass and fragment masses aligned for each lipid identified.
Creating a report
LipidView software enables users to create a report directly to MS Word format. Samples and processing methods are automatically printed on the report template and users can create numeric tables automatically from the results workspace. Additionally, lipid profile plots and labeled spectral views can be sent to the report directly from the results workspace, Advanced Profiles View, or Spectral View. Reports generated are automatically saved with a given LipidView software session and can be easily re-accessed when reviewing archived lipid result.
Sharing lipid details across SCIEX software
The processing methods of LipidView software hold the information on the studied set of lipid species, as well as their respective parent and fragment m/z values and internal standards. This information can be converted to formats used in other SCIEX software with the LipidView method exporter tool. Three types of export text formats are available: one can be imported as a MultiQuant software processing method, the other one can be imported into MasterView software and a textual lipid MRM list that can be pasted into the MS acquisition method interface in Analyst software. The export types and additional export parameters are shown in Figure 7.
LipidView software enables processing and review of MS and MS/MS data from infusion and LC ESI experiments from a variety of different acquisition types – supporting Analyst® Software raw data files.