| Date: | 03/02/2026 |
| Categories: |
For research use only. Not for use in diagnostic procedures.
Issue Description
After launching Analyst® software 1.7, the following error message appears:
Feature has expired
Feature: ANALYST.LAUNCH
Expire date: 15-jun-2020
License path: C:\ProgramData\ABSCIEX\Analyst\License\Analyst1.7.lic;
Flexnet licensing error:-10,32
After clicking OK, the error popup screen closes. There is no prompt for a new license code, and the software closes.
Resolution
This error message indicates that the Analyst™ software 1.7 license has expired. To resolve the issue, the expired trial license file needs to be deleted from the directory, C:\ProgramData\ABSCIEX\Analyst\License. To launch Analyst software again, replace the trial license with a permanent license or a new, unexpired trial license.
Posted: now