Audit records from Analyst® software accumulate in the Project Audit Trail and Instrument Audit Trail, and these files can create large files that are difficult to navigate and manage. Quantitation Audit Trails typically have a smaller, more manageable number of records.
When the Instrument Audit Trail or a Project Audit Trail reaches 1000 records, a final record stating that the file has been archived is added. The audit trail is automatically saved in the Project Information folder with a name indicating the type of audit trail and the date and time (for example, “PAT-Archive-200209300820.ata”). A new Instrument Audit Trail or Project Audit Trail is created, and the first record of the archived audit trail gives the path.