Answer

BioPharmaView™ Software 3.x and BioPharmaView (BPV) Flex Software 2.x are compatible and can be installed on the same computer; however, data files created using each software are not co-compatible and can only be opened or further processed using the originating software.

The two versions of software are each built using different software architecture, such that project files created in BioPharmaView Software 3.0 cannot be opened in BPV Flex. The only option for opening BioPharamaView software files in BPV Flex software is to create new projects in BPV Flex and then reprocess those data sets in BPV Flex.

Note that results from the two processing workflows, IP and PM, will be different between BioPharmaView™ Software 3.0 and BPV Flex, due to improvements made to the processing algorithms in BPV Flex.