Can I display the pressure trace of a chromatogram when I use an Agilent LC system with Analyst


Date: 06/02/2016
Categories: Analyst software

   Print    Rate Article:

For research use only. Not for use in diagnostic procedures.
Answer 

The pressure trace has to be activated in the hardware profile.

With deactivated profile, go to “Edit Profile”, click on the pump, then on “Setup Device”, and activate the checkbox “Enable Real-Time Monitoring”

When you acquire data with this hardware profile, the pump pressure can be displayed in Analyst Explore mode with right mouse click “Show ADC data” (in PeakView software: Show – UV/DAD/ADC channel)