Error Message for National Instruments GPIB Driver Received during Analyst® Software Installation


Date: 03/02/2026
Categories: Analyst software

   Print    Rate Article:

For research use only. Not for use in diagnostic procedures.


Issue Description

While installing Analyst® software, the following error message was received:
"National Instruments GPIB Driver
:The following driver is required for data acquisition and is not present on the system. This driver is not required for systems that will be used for data processing only or for those systems using an Ethernet connection."
User-added image

 


Resolution

Upon completion of the Analyst software installation, browse to the drivers folder on the DVD. In the GPIB folder, click Set up.exe and then follow the instructions in the installation wizard to install the driver.

Ensure that Analyst software and the driver are being installed from a local admin Windows account.