How Do I Change the Significant Figures of Regression Equation in Analyst


Date: 10/01/2015
Categories: Analyst software

   Print    Rate Article:

For research use only. Not for use in diagnostic procedures.

Answer 

The following procedure must be carried out by a User with administrative permissions on the computer.

  1. Close any open results tables.
  2. Go to start menu of windows and type ‘regedit’ into the search bar.  When regedit.exe program is found open it.
  3. Navigate through the tree view to find HKey_Current_User > PE Sciex > Analyst > Explore > Regression.
  4. Double Click on Num Decimals to change the value.
  5. Enter the required number of decimals and click OK.