How to export the Sample type in automatic reports generated by ChemoView™


Date: 03/02/2026
Categories: Analyst software , ChemoView software

   Print    Rate Article:

For research use only. Not for use in diagnostic procedures.


Answer

ChemoView™ cannot use all of the avaliable Analyst® tags in report templates. In ChemoView™ report templates most of the tags use tag called CUSTOM :

  • The Sample_type tag does not work in Chemoview reports. No values will be reported.
image.png
  • The Custom Sample_type tag does not work in Chemoview reports. The Sample ID value is visible instead of the value of the Sample type.
image.png

The workaround is to export the Sample type information in the report generated by ChemoView :

1) The first step for this workaround is to enter the Sample type in the Sample ID column within the Analyst® batch.

image.png

2) The final step is to use the Custom Sample type tag in the Chemoview report template as described above.


The limitation of this workaround : It is not possible to export the Sample type and the Sample ID simultaneously.