Formatting Analyte Names in an Analyst® Software Quantitation Method


Date: 03/02/2026
Categories: Analyst software

   Print    Rate Article:

For research use only. Not for use in diagnostic procedures.


Issue Description

When running a report in Analyst® software, the template doesn't populate results summary tables when the quantitation method is changed or updated.

The report shows the column headings, but there is no data populating the columns, as shown below:

The report should look like the image below:
Good report which shows data.


Resolution

If the dashes are used within an analyte name in the quantitation method, the results will not be populated correctly on the report. Use a space to separate analyte name and number in quantitation method to correctly group analytes

Below is a screenshot that shows how a quantitation method should be set up (without dashes and using space to separate analyte name and number).
User-added image

Below is a screenshot of a quantitation method that will not show results in the report because it uses dashes. 
User-added image