How Is the ITC Parameter Accessed in Analyst® TF Software?


Date: 03/02/2026
Categories: Analyst software , QTOF systems

   Print    Rate Article:

For research use only. Not for use in diagnostic procedures.


Answer

Changing the ITC parameter in Analyst® TF Software helps to assess if the QJet® ion guide on a Triple TOF® system needs to be cleaned. This article explains how to access the ITC parameter from the Analyst software. (Images are of Analyst TF 1.7.1 Software on a computer with a Windows 10 operating system installed.)

Step 1: Activate a hardware profile.
User-added image

Step 2: While the Configure mode is still active, go to Tools and then Settings and select Parameter Settings.
User-added image

Step 3: Select the proper scan type (generally TOF-MS +) and then double click on the ITC parameter to open it up. It is recommended to take a screen shot of the current settings and save them for future reference. These settings will need to be returned to their original configuration later.
User-added image

Step 4: Enter the value you wish to test and hit enter. (In this example, 0 is changed to 1.)
User-added image

Alternate Step 4: Change this parameter form a Fixed parameter to an Operator parameter by selecting Operator on the screen below.
User-added image

The Parameter Settings Editor screen would appear as follows:
User-added image

Now, when a new method is built, the ITC is available in the Parameters section of your method.
User-added image

Step 5: When the testing is complete, it is very important to change the parameter settings back to the default for your instrument. In particular, the ITC parameter needs to be set back to the original settings found in Step 3. The ITC should be to be set to 0 for general use to allow Analyst TF to automatically adjust this parameter as needed.