Sciex OS 1.5 Software Error: Processing Method Contains No Components or Is Incompatible


Date: 08/09/2019
Categories: SCIEX OS

0 Votes
   Print    Rate Article:

For research use only. Not for use in diagnostic procedures.


Issue Description

When opening a processing method in Sciex OS 1.5 software, an error was received, indicating that the method contained no components or was incompatible. The processing method that generated the error could still be used post-acquisition to process data, but it could not be assigned at the batch submission level. 

User-added image


Resolution

To resolve this error message, remove any subfolders in the project folder The user had created a sub-folder at the following level: C:\SCIEX OS Data\<PROJECT NAME>\Quantitation Methods\<Sub-Folder>. Removing the sub-folder resulted in normal software operation. 

Further, this issue is reported to be fixed in Sciex OS 1.6.1.
Â