DiscoveryQuant™ Software Produces SQL Error during Migration from Windows 7 to Windows 10


Date: 03/02/2026
Categories: DiscoveryQuant software

   Print    Rate Article:

When DiscoveryQuant™ software was installed on Windows 7, the SQL database associated with the software was using a UDL file that had a username but no password; this was the cause of the error. 

To resolve the error, a second user account on the SQL server was created with a password enabled. Using the new account and password allowed for a successful connection with DiscoveryQuant software. It is recommended not to use SQL user accounts without a password enabled on Windows 10.