Date: | 08/04/2023 |
Categories: | Software |
For research use only. Not for use in diagnostic procedures.
Answer
When making an Exion LC 2.0 diode array (DAD or DAD-HS) or multiwavelength detector (MWD) method in SCIEX OS software, a sampling rate of only 10 Hz or lower is supported. An LC method with a sampling rate greater than 10 Hz will not be saved.
Posted: now