Kay Hup1, Abdessamad Chahbouni1, Jack Steed2, Bertram Nieland3, Said El Ouadi3, Daniel McMillan2 and Jianru Stahl-Zeng4
1Het Waterlaboratorium; 2SCIEX, UK; 3SCIEX, The Netherlands; 4SCIEX, Germany
This technical note demonstrates the robustness of the SCIEX 7500 system over 15 months of routine PFAS analysis with various water samples. Quality control (QC) samples, spiked at 10 ng/L with L-PFHxS, L-PFOS, L-PFOA and L-PFNA, showed accuracies generally within ±1 standard deviation of the mean and all QC samples were with 30% of the mean values (Figure 1). Only 1 preventative maintenance (PM) service was performed during the 15-month timeframe. During the analysis, approximately 100–200 injections were performed per week and other non-PFAS applications were also run. These results highlight the strong robustness of both the analysis method and the SCIEX 7500 system.
A robust analysis is paramount to the long-term viability of routine PFAS analysis. In this technical note, the developed method was validated for 26 PFAS compounds of interest in LC-MS-grade, drinking, ground and surface water samples.1 Modifications were made to the LC system to reduce background contamination, including replacing or bypassing any components of the system that contribute to PFAS contamination. Specifically, system components containing fluorinated ethylene propylene (FEP) and Teflon were bypassed or replaced with polyether ether ketone (PEEK), when possible.
To further clarify how the QC data were collected, the analysis batch is included below. Step 8 introduces the QC samples analyzed in matrix. Step 14 outlines the flushing procedure used following sample analysis.
Prior to sample analysis, the consumables used for sample preparation and storage were flushed with 50:50 (v/v), methanol/acetonitrile to ensure that background levels were minimized. These consumables included the sample centrifuge tubes with screw cap and the polyethylene (PE) syringe for filtration. Resistor capacitor (RC) filters were flushed and the first 2 mL of filtrate and pipette tips (both positive displacement and air displacement tips) were disposed of.
To achieve clean blanks, modifications were made to the LC system to mitigate PFAS contamination as much as possible. The LC system used here was a Shimadzu LC-40 system. The changes described here might vary between systems, depending on configuration and flow path.
The original R-0 to R-2 rinse selection block in the autosampler was bypassed. Only 1 rinse liquid was used, which was connected directly to the low pressure valve (LPV) using a PEEK tube. Subsequently, the original Teflon 1/16” tubing connecting the LPV to the bottom of the rinse port was replaced by 1/16” PEEK tubing.
The LC parts replaced with PEEK included:
For complete method details and more information please refer to reference 1.