Methods
Material and solutions: Flow rate and injection precision, as well as injection linearity, of the SCIEX ExionLC 2.0+ system, were investigated using a Supelco HPLC Gradient System Diagnostic Mix (P/N 4-8271) containing uracil (dead volume retention marker), phenol, methyl paraben, ethyl paraben, propyl paraben, butyl paraben and heptyl paraben.
Chromatography: LC separations were performed using the SCIEX ExionLC 2.0+ system. For injection linearity and precision determination, a 2.6 μm Phenomenex Kinetex C18 column (2.1 x 50 mm, P/N: 00B-4462-AN) was chosen, and a simple gradient of water and acetonitrile, containing 0.1% formic acid, was used. The syringe speed was set to low and the speed factor to 0.1. For determining the flow rate precision at high pressure, an ACQUITY UPLC HSS T3 1.8 µm 2.1 x 50 mm was selected. Mobile phase was water/acetonitrile. The LC flow rate was 1.2 mL/min and the LC run-time was 3.25 minutes, with a maximum pressure of 1,100 bar (16,000 psi). The 1.8 µm particle column was replaced with a Phenomenex Synergi 4 µm Fusion-RP 2 x 50 mm column (PN 00B-4424-B0) to investigate the low-pressure method performance on this high-pressure system with a minimum pressure of 110 bar (1,600 psi). The SCIEX ExionLC 2.0+ system autosampler was used with a configuration consisting of a 250 µL syringe, 250 µL buffer tubing and 15 µL needle tubing. A 10 µL sample loop was used to test the partial loop injection mode with low injection volume injections. In order to optimize sample consumption and minimize the injection cycle time, the µL pick-up plus injection mode was selected for the 100 injections retention time and area reproducibility experiments, which required using a 100 µL sample loop.
Diode array detection: Ultraviolet-visible detection was performed using an ExionLC 2.0+ system with integrated ExionLC 2.0 diode array detector HS, equipped with a 10 mm, 10 μL, 300 bar flow cell. For the Supelco HPLC Gradient System Diagnostic Mix, the detector was operated at 254 nm. All diode array detector data collection was performed with a data rate of 10 Hz.
Mass spectrometry: Data acquisition was performed using Analyst software 1.7.1 with Components for the ExionLC 2.0+ system. It is worth noting that the ExionLC 2.0+ system is also fully supported for instruments in which data acquisition is performed using SCIEX OS software.
Data processing: Data processing of diode array detector acquired data was performed using SCIEX OS software 2.0.1 in which calibration curves, precision and accuracy statistics were generated.
Flow rate precision
Figure 1 shows overlaid chromatograms from 100 consecutive fast, 78 second gradient separations and the retention time stability of the SCIEX ExionLC 2.0+ system running with a pressure of 1100 bar. As shown in Figure 2, the retention time precision of each of the analytes eluting at different stages of the gradient for these injections is less than 0.15% RSD. For compounds tested, the maximum retention time difference over 100 injections was less than 0.3 seconds.
The flow rate precision experiment was repeated using the same gradient but replacing the 1.8 µm particle column with a Phenomenex Synergi 4 µm Fusion-RP 2 x 50 mm column to produce lower pressure conditions. This was to allow investigation of the low-pressure method performance on this high-pressure system, running with a minimum pressure of 110 bar (1,600 psi).
Figure 3 shows overlaid chromatograms from 100 consecutive injections and the retention time stability obtained from the SCIEX ExionLC 2.0+ system running at conventional, lower pressure conditions [110 bar(1,600 psi)].
Table 1 shows retention time variability results for these 100 injections, and the flow rate precision of the SCIEX ExionLC 2.0+ system when running at 110 bar (1,600 psi). The retention time precision of each of the analytes eluting at different stages of the gradient for these injections is ≤ 0.12% RSD. For compounds tested, the maximum retention time difference over 100 injections was less than 0.3 seconds.
The results show that there is no detrimental effect on LC pump stability with respect to retention time when using the SCIEX ExionLC 2.0+ system at pressures as low as 110 bar. This demonstrates that the SCIEX ExionLC 2.0+ system can be used for both ultrahigh pressure liquid chromatography workflows, using sub 2 µm column particle packing, and conventional low-pressure methods with columns of larger particle size.
Utilizing the µL pickup plus injection mode in these experiments ensures no excess sample is used. The injection sequence has been optimized so that the time to injection is not compromised over other injection modes which can be made in less than 17 seconds.
Injection linearity and precision
A mixture of compounds was analyzed using the integrated ExionLC 2.0 diode array detector HS for detection. Very good separation of the components was achieved (Figure 4). Replicate injections (n=5) were performed for small injection volumes between 0.5 and 5 µL to investigate the use of a 10 µL sample loop and partial loop injection mode. The area counts were measured and plotted against corresponding injection volumes for each component in the Supelco HPLC Gradient System Diagnostic Mix.
Figure 5 shows that the linear correlation coefficient (r) regression analysis for all compounds is > 0.999.
The injection reproducibility was also computed from these same data, for all the various injection volumes. Very low variance was observed, even for the very low injection volumes. Figure 6 shows that the %CVs for all injection volumes are below 1%.
Conclusions
The ExionLC 2.0+ system is a flexible and robust UHPLC system that is suitable for today’s challenging LC-MS/MS workflows.
- Retention time precision is critical when running multi-analyte panels and time scheduled MRM assays. High flow rate precision of the ExionLC 2.0+ system, using fast gradient separations (sub 90 seconds), provided RSD of <0.15% for the analytes tested here.
- With the high pressure ExionLC 2.0+ systems, high-performance LC-MS workflows can be achieved with both high and low-pressure applications.
- High autosampler precision and accuracy are also essential for quantitative experiments. As demonstrated here, this system can achieve <1% CV across the full range of injection volumes tested (0.5 –5 µL) using a 10 µL sample loop and partial loop injection mode. Excellent linearity was also demonstrated (linear correlation coefficient (r) regression analysis was > 0.999 for all analytes).